Bio Medical Solution

Products

Bio Safety Cabinet Class II Type A2

Click To Scan

Bio Safety Cabinet Class II Type A2

Description

Biological Safety Cabinet Class-II Type-A2 designed to protect the Laboratory staff, environment and sensitive work processes in which biological agents are used. The safety cabinet offers a superior level of contamination protection based on two advanced HEPA filters. The Biosafety cabinet configuration is freestanding. The Work zone and side walls are made of one piece 16gauge stainless steel 304 grade 4B finish material.The Cabinet body is made of of  18gauge electro-galvanized steel with oven baked epoxy polyester anti-microbial powder coating material. In order to ensure consistent and reliable down flow velocity across the supply HEPA/ULPA filter over the life of the cabinet, the cabinet use a pressure sensor (rather than anemometer) to detect pressure drop across the supply filter, rather than just one point across the down flow. The pressure sensor encased in order to protect the sensor from temperature, humidity, and other environmental phenomena that can impact the sensor’s performance. In order to preserve the safety of the user and the environment, the exhaust blower on the cabinet must continue operating when the supply blower stops working. If the exhaust blower  fail, the supply filter will also be turned off. The Cabinet motors automatically adjust the airflow speed without the use of a damper to ensure continuous safe working conditions, even without maintenance adjustments.

Features

  • ISO 14644 – 1 class 3 air quality
  • Incorporate HEPA filter of the class H14 EN 1822
  • Heat dissipation 619 BTU/hr.
  • One brushless direct current electronically commutated motor
  • Gauges for monitoring the condition of all HEPA filters as well as work space
  • Microprocessor control panel with displayed for airflow velocity, filter life, UV/blower hours
  • Vertical airflow direction
  • 220-240V,50Hz with single phase AC power supply
  • Coved corners in the work zone and smooth drain pan edges allow quick and efficient cleaning
  • Quick start mode for fast operation
  • Power Consumption 197W
  • A one-piece spill-retaining work tray with large handles for convenient lifting
  • Audible and visual alarms for HEPA filter failure, blower failure, airflow speed failure and Incorrect window position
  • 7’’ touchscreen display
  • UV lamp with safety interlock operates only when sash is closed
  • A tray holder keeps the work tray firmly in place for convenient cleaning of the drain pan
  • Control panel positioned at the center with a downward tilt for improved visibility and operation
  • Optional gas/vacuum service fixtures
  • Low heat and low energy consumption 70% more energy savings
  • In order to provide maximum effectiveness, efficiency and safety to laboratory personnel, UV light programmable to allow for specific exposure times from 0 to 24 hours with separate switch to operate the UV light and automatic shut-off feature present for UV light
  • Cabinet usable work area 0.6 m2(6.5 ft2)
  • Electrical sockets are installed on the side panels for quick and easy reach
  • Dual layer side walls with negative pressure
  • Easy reach & viewing for centered and angled down
  • Support stand equipped with a combination of caster wheels and levelling feet
  • Touch-sensitive areas that support finger gestures
  • Cabinet automatically blower speed to 30% when the front window sash is in closed position to ensure reduced energy consumption when the cabinet is not is use
  • All electrical components are positioned behind the front panel
  • Paper catche removable
  • Silver-ion infused antimicrobial powder coating helps inhibit microbial growth, preventing the plenum from turning into a biohazard waste zone
  • Option to connect with Wi-Fi
  • Drain pan made of stainless-steel material
Specs/Model MBMSJ-1018MS-03 MBMSJ-1018MS-04 MBMSJ-1018MS-06
External Size (WxDxH)
1000 X 800 X 1500 mm
1300 X 800 X 1500 mm
1870 X 800 X 1500 mm
Internal Size (WxDxH)
900 X 600 X 600 mm
1200 X 600 X 600 mm
1700 X 600 X 600 mm
Sash Opening
254 mm
254 mm
254 mm
Air Flow Volume (Inflow)
350 CFM
350 CFM
350 CFM
Air Flow Volume (Downflow)
225 CFM
225 CFM
225 CFM
Air Flow Volume (Exhaust)
350 CFM
350 CFM
350 CFM
Average Down flow Velocity
60FPM
60FPM
60FPM
Average Inflow Velocity
105FPM
105FPM
105FPM
Airflow Pattern
70% re-circulation, 30% air exhaust
Front sash
Safety glass counter balanced with marked safe operating height
Supply/Exhaust Filter
ULPA filter efficiency 99.999% for 0.1μm particles
Noise Level
60 dB(A)
LED lamp illumination
1350 Lux
UV lamp
254 nm

Additional Accessories

  • Armrest
  • Gas burner  
  • Datalogger